GoodKnight 420 Series 420Sマニュアル 取扱説明書

AUTO CPAP Puritan Bennett Goodknight 420S Manuals:
患者さん用マニュアル取扱説明書
http://www.cpapaustralia.com.au/resources/file/GoodKnight%20420S%20Patient%20Manual.pdf

http://www.nbngroup.com/manuals/machine/420S.pdf
http://www.nbngroup.com/manuals/machine/420G.pdf

医療従事者用マニュアル取扱説明書
マニュアルはなし 設定のみ
Goodknight 420S Clinician Manual
Puritan Bennett GoodKnight 420 Series: Make sure the CPAP machine is plugged in, then find the secret button beneath the "Go" in the word GoodKnight which appears just above the LCD screen. Press the secret button and the Information Access button simultaneously for 2 or 3 seconds. The CPAP machine will now be in the provider setup mode. Press the Information Access button to cycle through the list of adjustable settings, and use the arrow buttons to make adjustments.

設定変更方法
GoodKnightのマークのところに隠れたボタンがある。(G「o」odKnightのところ)
そのボタンとインフォメーションボタン(3つのボタンの一番左)を同時に2、3秒間押す。
設定変更できるようになる。
インフォメーションボタンを押して該当項目を変更。
上ボタン(一番右)下ボタン(真ん中)を押して、レベルを調整する。
シンプルで使いやすい機器。
 
参考
http://www.apneaboard.com/adjust-cpap-pressure/change-cpap-pressure-settings-adjusting-your-machine-with-a-clinician-setup-manual

Puritan Bennett Manuals:
Puritan Bennett Sandman (Auto, Intro & Info)
Puritan Bennett Sandman Duo & Duo ST (combined manual)
Puritan Bennett GoodKnight 420g
Puritan Bennett GoodKnight 420e APAP
Puritan Bennett 425 Bilevel
Puritan Bennett KnightStar 330 Bi-Level
Puritan Bennett GoodKnight 418A and 418P (combined manual)
http://www.apneaboard.com/adjust-cpap-pressure/change-cpap-pressure-settings-adjusting-your-machine-with-a-clinician-setup-manual

I have not been able to find the clinical manual for 420S
Here is the patient manual http://www.nbngroup.com/manuals/machine/420S.pdf

Puritan Bennett GoodKnight 420 Series
To get into the clinician's setup mode in a 420 series CPAP machine follow these steps:

1- make sure the CPAP machine is plugged in
2- find the secret button beneath the two "o"s in the word GoodKnight which appears just above the LCD screen
3- press the secret button and the Information Access button simultaneously for 2 or 3 seconds.

The CPAP machine will now be in the provider setup mode. Press the Information Access button to cycle through the list of adjustable settings, and use the arrow buttons to make adjustments. Below is a list of settings you can cycle through on a Puritan Bennett 420G:

CPAP Pressure (only in provider setup mode)
Ramp Time
Ramp Start Pressure
Hour Meter Value (which cannot be adjusted)
Compliance Meter Value (which can only be reset to 0)
Offset of Pressure Sensor (used for calibration purposes) (only in provider setup mode)
Machine ID Number (which cannot be adjusted)